Tips for Medical Device Fatigue Testing

Calvin Chao 640 x 480

By Engaged Expert

Calvin Chao

More content from this author

Learn about our Engaged Experts

Calvin Chao is an expert in mechanical testing of cardiovascular devices, acute particulate testing and nickel ion leaching measurements.

Fatigue testing is utilized for characterizing material properties and specimen behavior during cyclic loading. Understanding the durability and limits of the device allows medical device manufacturers to bring safe and reliable products to market.

Given the complex regulation and high risk of medical device applications, fatigue testing must be stringently performed to the appropriate stress levels, sample size, and standards according to well-defined test protocols.

 

How is medical device S/N fatigue testing performed?

Medical device fatigue testing focuses on the nominal stress required for failure after a number of cycles. Stress (S) is typically plotted on a log scale against the number of cycles to failure (N), also known as S/N Fatigue Testing. 

During an S/N fatigue test, the first sample group is tested at high peak stress (or strain in a variation of the test) where failure occurs after a low number of cycles. Subsequent samples are tested at decreasing stress levels until a statistically significant number of samples do not fail in the specified number of cycles. The fatigue threshold is the highest stress at which a runout (non-failure) occurs.

 

How to Determine Sample Size 

Determining the appropriate sample size is a balance between the number of tests needed for the highest level of statistical confidence and the cost of developing the test specimens. The number of samples needed depends on the standard deviation and desired levels of statistical significance. Power analysis calculators are available online to help you determine the best sample size for your application. 

 

Samples and Runouts for Orthopedic Device Fatigue Testing

Orthopedic devices experience various forces during their use inside the body, including fatigue, torsion and wear. The number of samples needed for an orthopedic device fatigue test may vary based on device type, materials and more. 

During bone plate fatigue testing, for example, six samples are often used. Four of the samples are tested to failure at the higher load levels, and two are used as runouts. If there are minor changes made to the product or process, only the runout samples may need to be retested to confirm that the changes did not impact device performance.

 

Sample design and failure detection methods for Intravascular Device Fatigue Testing

For intravascular device tests without a predicate, the required sample count and number of test levels are often higher than with an orthopedic device. For example, a strutted stent design might use a six or twelve-specimen fixture for each test level.

To achieve better granularity on the cycles (N) and stress (S) axes, six or more test levels might be used with each six- or twelve-specimen test setup. While the number of tests might be six, the number of specimens tested could be 72. When testing multiple specimens, it is useful to incorporate a failure detection mechanism so that specimen failures can be tracked without the need for constant visual inspection.

Test coupons for intravascular device fatigue tests are often diamond-shaped for stents and valves and might be either cut directly from a device (1:1 scale) or scaled up, so they are easier to handle if device dimensions are very small.

 

The ÌÇÐÄlogoÃ×·ÆÍÃadvantage

As a comprehensive medical device testing partner, you’ll enjoy the benefit of a single source supplier for all of your testing needs, from product qualification and mechanical testing to material characterization and microbiological evaluation. 

Our medical device experts work diligently to help your company mitigate risk and bring safe products to patients worldwide. Contact us to discuss how we can help with your medical device fatigue testing. 

How to create a test protocol for medical device fatigue testing projects

To ensure a successful fatigue test project of any size, it is critical to perform the testing in line with a clear and thorough test protocol. Engaged Expert Maciej Jakucki outlines how to structure your document, analyze requirements and define acceptance criteria in our Test Protocol for Medical Devices article. 

 

Which medical device fatigue test system should I use?

There are different systems available for fatigue testing of medical devices.  While most 21st-century fatigue testing applications leverage linear motor technology, each technology possesses unique characteristics that make it indispensable to medical device testing. Our Comparison of Medical Device Fatigue Test Systems article explains the difference between Servohydraulic, Servopneumatic, Single Phase and Multi-Phase Linear Motor test systems. 

 

Do’s and Dont’s of Medical Device Mechanical Testing

For more guidance on fatigue testing and mechanical test programs for medical devices, read our article on the Do's and Don'ts of Medical Device Mechanical Testing. 